• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97702
Device Problems High impedance (1291); Device Difficult to Program or Calibrate (1496)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/25/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
On (b)(6) 2021, (b)(4) (con, rep, hcp): the manufacturer representative (rep) reported that the patient had heard the surgeon and a manufacturer representative (rep) say the connections were bad on the date of implant, but everything would be fine.   on (b)(6) 2021, the rep was unable to program the patient.  impedances were checked and all impedances were over 40,000 ohms.  the patient was going to be referred to a neurosurgeon, but the rep confirmed surgery was not yet planned or scheduled.  no symptoms reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key12388497
MDR Text Key272907793
Report Number3004209178-2021-13103
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169974883
UDI-Public00643169974883
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/14/2021
Device Model Number97702
Device Catalogue Number97702
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/25/2021
Date Device Manufactured02/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age79 YR
Patient Weight106
-
-