Model Number 05.000.008
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional device product codes: gxl and hxx.Reporter is a synthes employee.Part #: 05.000.008.Synthes lot #: 006253.Supplier lot #: n/a.Release to warehouse date: june 1, 2018.Manufactured by: synthes monument.No ncr's generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.The repair technician reported the device did not run in forward and found some sort of sulfation on motor.Damaged component is the reason for repair.The cause of the issue is damaged component.The following parts were replaced: circuit board, motor, membrane switch/flex circuit, and all applicable components.The item was repaired per the inspection sheet, passed synthes final inspection and will be returned to the customer upon completion of the service and repair process.Finalized service record will be archived in tungsten document management system.The evaluation was confirmed.The device was deemed serviceable and will be returned to the customer, no design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date during a routine field maintenance, the battery powered drivers were found to not be functioning properly.There was no patient involvement.This report involves 1 hand piece for battery powered driver.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAND PIECE FOR BATTERY POWERED DRIVER
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12391001
MDR Text Key268904731
Report Number2939274-2021-05036
Device Sequence Number1
Product Code HWE
UDI-Device Identifier10887587024585
UDI-Public(01)10887587024585
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number05.000.008
Device Catalogue Number05.000.008
Device Lot Number006253
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2021
Date Manufacturer Received08/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-