• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM Back to Search Results
Model Number 283512
Device Problems Mechanical Jam (2983); Mechanics Altered (2984)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4). Investigation summary: the device was received at the service center and evaluated. There was no allegation of malfunction against the device from the customer, however, defects were found with the device during service evaluation. This complaint can be confirmed. During the service evaluation the following defects were identified: the motor shaft is stuck - functional : will not run motor(s). The locking ring does not close properly. The defective motor and defective drill mounting mechanism 1. 0 were replaced to resolve the issues. After repair, the device was found to be working according to the specifications. The faulty parts was identified as the root cause for the device failure during the service evaluation. Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing. At this point in time, no corrective action is required, and no further action is warranted. Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the affiliate in (b)(6) that during service evaluation, it was determined that the locking ring did not close properly on the micro tornado hp w handcontrol device. There was no procedure nor patient involvement reported. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMICRO TORNADO HP W HANDCONTROL
Type of DeviceLINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key12406674
MDR Text Key269606390
Report Number1221934-2021-02584
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 09/01/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2021
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/01/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown
Removal/Correction NumberN/A

-
-