• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. UNIPOLAR HIP; HEAD, UNIPOLAR, MODULAR, ENCORE SIZE47

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. UNIPOLAR HIP; HEAD, UNIPOLAR, MODULAR, ENCORE SIZE47 Back to Search Results
Model Number 411-00-147
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem Unspecified Infection (1930)
Event Date 08/18/2021
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 35 days apart.Initial or prolonged hospitalization was required.The healthcare professional indicated there was a serious risk to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post-surgical instructions.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Revision surgery: due to infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIPOLAR HIP
Type of Device
HEAD, UNIPOLAR, MODULAR, ENCORE SIZE47
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
kiersten soderman
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key12407411
MDR Text Key269362968
Report Number1644408-2021-00919
Device Sequence Number1
Product Code KWL
UDI-Device Identifier00888912076821
UDI-Public(01)00888912076821
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/02/2023
Device Model Number411-00-147
Device Catalogue Number411-00-147
Device Lot Number914B1047
Was Device Available for Evaluation? No
Date Manufacturer Received08/18/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/12/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
411-00-035 LOT 903B2051
Patient Outcome(s) Hospitalization; Other; Required Intervention;
Patient Age78 YR
-
-