MAUDE Adverse Event Report: FLOWONIX MEDICAL, INC. PROMETRA II PROGRAMMABLE PUMP; IMPLANTABLE INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FLOWONIX MEDICAL, INC. PROMETRA II PROGRAMMABLE PUMP; IMPLANTABLE INFUSION PUMP Back to Search Results
Model Number 13827
Device Problems Communication or Transmission Problem (2896); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/09/2021
Event Type  Injury  
Manufacturer Narrative
A review of the device history record, which includes verification of all steps in the manufacturing of the pump, verification of all final testing performed by/on the pump, verification of sterilization, and packaging for subject pump was performed.The review did not identify any non-conformances, issues or capas associated with pump function.Device remains implanted and was not returned.Physician did not know a cause for the pump flipping.Patient was not manually flipping the pump.Per the instructions for use of the device, pump flipping is a known possible risk of use of the device.Internal complaint number: (b)(4).
 
Event Description
Clinical specialist (cs) reports that the patient's pump had loosened in the pocket and that it was flipping.It was confirmed that the pump was initially sutured down.The patient was scheduled for a refill and they were unable to inquire it so the physician scheduled the revision.Cs reported that a pocket revision was performed to move the pump to the patient's back.Cs also reported that the surgeon decided to replace the catheter because it had become loose from the anchor and was wrapped around the pump.Cs reported there was no damage or adverse effects to the patient.The doctor did not say why they believed that the pump was flipping.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMETRA II PROGRAMMABLE PUMP
Type of Device
IMPLANTABLE INFUSION PUMP
Manufacturer (Section D)
FLOWONIX MEDICAL, INC.
500 international drive
suite 200
mount olive NJ 07828
Manufacturer (Section G)
FLOWONIX MEDICAL, INC.
120 forbes blvd
suite 170
mansfield MA 02048
Manufacturer Contact
james bennett
500 international drive
suite 200
mount olive, NJ 07828
9734269229
MDR Report Key12408437
MDR Text Key269443356
Report Number3010079947-2021-00253
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00810335020228
UDI-Public(01)00810335020228(17)220218(10)27525
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/18/2022
Device Model Number13827
Device Catalogue Number13827
Device Lot Number27525
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/09/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-