Catalog Number UNKNOWN
Device Problems Structural Problem (2506); Device Tipped Over (2589)
Patient Problems Internal Organ Perforation (1987); Perforation of Vessels (2135); Anxiety (2328)
Event Type  Injury  
Manufacturer Narrative
Non-healthcare professional.Investigation: the reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Catalog number and lot number are unknown; however, there is no evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.The following allegations have been investigated: organ/vena cava (vc)/aorta perforation, tilt.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
 
Event Description
The following information is alleged: the patient received a celect pt inferior vena cava (ivc) filter on (b)(6) 2017 and several prongs of the celect platinum filter have perforated the ivc.Struts further perforate the aorta and duodenum.The filter has also tilted substantially.Hospital and medical records have been requested, but not yet provided.
 
Manufacturer Narrative
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: anxiety, worry, physical limitations.Unknown if the reported anxiety, worry, and physical limitations are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Patient allegedly received an implant on (b)(6) 2017 via the right internal jugular vein due to upcoming surgery.Patient is alleging tilt, vena cava/organ perforation, and organ abutment.The patient further alleges anxiety, worry and physical limitations.Report from ct (computed tomography): "ivc filter is present within the infrarenal ivc.There is filter tilt with the superior aspect of the filter adjacent to the lateral wall of the ivc.Several struts course through the caval lumen with one strut within the wall of the infrarenal abdominal aorta extending a distance of approximately 3 mm into the aorta.To [sic] additional struts is adjacent to the wall of the third portion of the duodenum and likely contact the wall.One of the struts is within the inferior aspect of the left renal vein.No other abnormality identified of the ivc filter.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
lissi walmann
sandet 6, dk-4632
bjaeverskov 
56868686
MDR Report Key12409097
MDR Text Key269585137
Report Number3002808486-2021-01762
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient SexMale
Patient Weight88 KG
Patient RaceBlack Or African American
-
-