DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3013-53 |
Device Problem
Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device (10/213) a getinge field service engineer (fse) evaluated the iabp unit and found that there were 2 alarms that put the unit into standby the "no trigger" and "check iab catheter".The fse let the device run with a test balloon and system trainer for an hour without the device going into standby and then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) kept pausing and going into standby.The customer requested that the unit be checked out for proper operation.No patient harm, serious injury or adverse event was reported.
|
|