Model Number 10-85-00 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova deutschland manufactures the s5 double head pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.The motor control board and the motor power amplifier board were replaced but the issue persisted.Thus, the failure was most likely due to a defective pump head which has been put on order and will be replaced.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova received report that a s5 double roller pump gave a motor control failure error message during priming.There was no patient involvement.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: the pump head was replaced and the reported error message did not recur.Subsequent functional verification testing was completed without further issues.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H.10: the root cause of the event has been traced back to a faulty resolver in pump head.
|
|