H.10: the controller data log file was reviewed.The data log showed that the pump was operated using controller (b)(6).Data captured during this time indicated that the pump remained stable within allowable tolerance around the set speed of 6600rpm.Near the end time of the case, measured flow, estimated flow, and pump current all decreased while speed remained steady.This is consistent with a restriction in flow occurring outside of the pump.The lifesparc controller is designed to operate the pump at a continuous rotational speed selected by the operator.The controller indicates a measured flow if a flow sensor is connected and an estimated flow otherwise.Measured flow is computed by measuring transit time with an ultrasonic flow sensor clamped to the tubing.Estimated flow is computed based on pump speed and current.At a fixed speed, a higher current indicates greater work performed by the pump and therefore a higher flow.Clot formation on the moving parts of the pump can cause an increase in estimated flow without a corresponding increase in actual flow, since the added mechanical resistance caused by the clot can increase the required current.Clot formation outside of the moving parts of the pump can cause a decrease in actual flow which also appears as a decrease in estimated flow.Through follow up communication it was also reported that the low flow condition was possibly due to clots of unknown location present during support.Based on the above facts and considering the investigation results, a malfunction of the lifesparc controller can be ruled out.The most likely root cause of the event is a clots/throumbus obstruction possibly due to patient condition causing a decreasing flow and resulting in the reported event.
|