During a procedure it was intended to use one resolute onyx device to treat a mildly tortuous lesion in the distal right coronary artery (rca) with 100% stenosis.
No damage was noted to the packaging, no issues were noted when removing the device from the hoop/tray.
The device was inspected with no issues.
Negative prep was performed with no issues noted.
The device did not pass through a previously deployed stent.
It was reported that the device was dislodged, detached, cracked or fractured during delivery through the vessel.
The device was not kinked and re-straightened during use.
The detached portion of the device was retrieved from the patient.
The device was replaced by another medtronic product.
The patient is alive with no injury.
|