Model Number 20
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2021
Event Type  malfunction  
Manufacturer Narrative
Physio-control advised the customer that their device is not field-serviceable and is no longer under warranty. Physio recommended that the customer remove the device from service and a replacement device be obtained. The device was not returned to physio-control for evaluation. The cause of the reported issue could not be determined.
 
Event Description
A customer contacted stryker to report that their device gave an "abnormal energy delivery" message, which could result in wrong or inappropriate defibrillation therapy being delivered. There was no reports of patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of DeviceAUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key12433590
MDR Text Key272170077
Report Number0003015876-2021-01713
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/07/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number20
Device Catalogue Number99402-000087
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/11/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/18/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No