Model Number 11.1.1
Device Problem Application Program Problem: Parameter Calculation Error (1449)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/07/2021
Event Type  malfunction  
Event Description
Merge cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information from merge healthcare and other vendors systems including images, hemodynamic studies and reports, measurements (via import from dicom structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data. Merge cardio is intended to allow users to review diagnostic and non-diagnostic quality images, annotate studies, perform digital subtraction on images, to perform quantitative measurements on images (including but not limited to quantitative coronary analysis, left ventricular analysis, time, area, length, velocity, angle, volume, and velocity-time integrals), to generate physician generated clinical reports (via structure reporting and template based tools), and to store this information in a database. On (b)(6) 2021, a customer contacted merge healthcare to report that the mr max gradient measurement was mapped incorrectly. Merge healthcare technical support was able to work with the customer to remap the measurement and confirmed that it was displaying correctly. No further action is necessary at this time. This has the potential to delay patient treatment and/or diagnosis. There have been no reports of patient injury or harm as a result of this issue. Reference complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMERGE CARDIO
Type of DevicePICTURE ARCHIVING AND COMMUNICATIONS SYSTEMS AND WORKSTATIONS
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
brian bell
75 binney street
cambridge, MA 02142-1123
3123540189
MDR Report Key12436343
MDR Text Key277744882
Report Number2183926-2021-00029
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Type of Report Initial
Report Date 09/07/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number11.1.1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/24/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-