• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX M100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX M100; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955595
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/13/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone no.(b)(6).The device was not received for evaluation however photos were provided.The visual inspection observed a crack on the access luer connector cone.The reported condition was verified.The cause is supplier related.This issue is being further investigated.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during treatment with a prismaflex m100 set, air bubbles were observed due to a cracked red luer connector.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M100
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key12438150
MDR Text Key270280676
Report Number8010182-2021-00264
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414124397
UDI-Public(01)07332414124397(10)
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial
Report Date 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Catalogue Number955595
Device Lot Number21C2503
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/13/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-