It was reported to philips that the defibrillator is out of tolerance.The device was received by bench for repair.The noted failure was identified during inspection of the device by an authorized bench technician.As a result of the issue, it was determined that the defibrillator capacitor would need to be replaced.Diagnostic performed by engineer: diagnostics of mrx defibrillator, serial number:(b)(6) performed in accordance with the service manual.No reported malfunction.During the tests, a reduced defibrillation power was found.Remaining functional tests were performed and ended with positive results.The device remains out of order - the condenser needs to be replaced in order to repair the device.Upon conclusion of the evaluation, it was determined that this was a malfunction of the defibrillator capacitor.The defibrillator capacitor was replaced to resolve the reported issue and the device was scheduled to be returned to the customer site.No further evaluation is required at this time.
|