• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC HEARTSTART MRX MONITOR/DEFIB; DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC HEARTSTART MRX MONITOR/DEFIB; DEFIBRILLATOR Back to Search Results
Model Number M3535A
Device Problem Output below Specifications (3004)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2021
Event Type  malfunction  
Event Description
It was reported to philips that the defibrillator is out of tolerance.There was no patient involvement.
 
Event Description
It was reported to philips that the defibrillator is out of tolerance.The device was received by bench for repair.The noted failure was identified during inspection of the device by an authorized bench technician.As a result of the issue, it was determined that the defibrillator capacitor would need to be replaced.Diagnostic performed by engineer: diagnostics of mrx defibrillator, serial number:(b)(6) performed in accordance with the service manual.No reported malfunction.During the tests, a reduced defibrillation power was found.Remaining functional tests were performed and ended with positive results.The device remains out of order - the condenser needs to be replaced in order to repair the device.Upon conclusion of the evaluation, it was determined that this was a malfunction of the defibrillator capacitor.The defibrillator capacitor was replaced to resolve the reported issue and the device was scheduled to be returned to the customer site.No further evaluation is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX MONITOR/DEFIB
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
dana tackett
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key12443511
MDR Text Key270492406
Report Number3030677-2021-14193
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838000018
UDI-Public00884838000018
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3535A
Device Catalogue NumberM3535A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date08/18/2021
Date Manufacturer Received08/26/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-