MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS SYRINGE 1ML LL W/O DN PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS SYRINGE 1ML LL W/O DN PISTON SYRINGE Back to Search Results
Model Number 309628
Device Problems Volume Accuracy Problem (1675); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2021
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown. The date received by manufacturer has been used as a default. Medical device expiration date: unknown. Device manufacture date: unknown. Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident. No root cause can be determined as no samples were received. The lot number is unknown, therefore device history record review (dhr) or quality notification review (qn) could not be performed. Examination of the product involved may provide clarification as to the cause for the reported failure. Complaints received for this device and reported condition will continue to be tracked and trended. Information will be captured on trend reports and monitored monthly. Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that syringe 1ml ll w/o dn had a scale marking issue and was damaged, but still operable. The following information was provided by the initial reporter: "finger flanges on item 309628 are bent and the volume markings appear to be inaccurate. ".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYRINGE 1ML LL W/O DN
Type of DevicePISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12447552
MDR Text Key270619724
Report Number1213809-2021-00624
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K941562
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/30/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number309628
Device Catalogue Number309628
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/11/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-