A user facility biomedical technician (biomed) reported finding thermal damage on the l2 cable going from the motor protection switch (mps) to the power contactor on an aquabplus 1500.
Someone from the clinical staff noticed a yellow warning light on the reverse osmosis (ro) monitor on the treatment floor, and the biomed went in the back to further investigation.
The biomed discovered that the concentrate pressure was too low, and the aquabplus 1500 was giving the alarm code w-02-51-03.
Upon further inspection, the biomed found charring on the l2 cable at the mps.
All cables connecting the mps to the power contactor were loose, but the charred (or burnt) cable was the loosest.
The biomed disconnected power from stage 2 and then replaced the mps and the burnt cable.
The spade connectors on the cables were then tightened, and the stage 2 pump resumed function.
There was no patient injury or harm, and no missed treatments.
Treatments were not impacted during the repair because the ro continued to operate under stage 1.
There was no evidence of any burning smell, smoke, sparks, or flames.
The biomed confirmed there were no local power grid issues in the area around the time of the event.
The biomed checked the local power supply and confirmed that there were no blown fuses there.
It was also confirmed that the thermal overload switch was not tripping.
The suspected cause of the failure was the loose cables.
The loosest cable was the one that had thermal damage, and its suspected that arcing may have occurred at this connection.
A photo of the charred cable was provided for review, as were the machine data files.
The mps and burnt cable were not available to be returned for manufacturer evaluation as they were discarded.
|