OPTIFLOW + ADULT NASAL CANNULA; BTT Back to Search Results
Model Number OPT944
Device Problems Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).We are currently in the process of completing our investigation.We will provide a follow up report upon completion of investigation.
 
Event Description
A distributor reported on behalf of a healthcare facility in (b)(6) that the tubing of the opt944 optiflow + nasal cannula was found damaged during patient use.There were no reported patient consequences.
 
Event Description
A distributor reported on behalf of a healthcare facility in china that the tubing of the opt944 optiflow + adult nasal cannula was found damaged during patient use.It was further reported that the patient tried to remove the subject opt944 optiflow + adult nasal.There were no reported patient consequences.
 
Manufacturer Narrative
(b)(4).The opt944 optiflow + adult nasal cannula is an interface used to deliver humidified oxygen to patients and consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Method: the complaint opt944 optiflow + adult nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the information and photograph provided by the customer, and our knowledge of the product.Results: visual inspection of the photograph revealed that the complaint cannula tubing was detached from the swivel connector.The healthcare facility further reported that the patient tried to remove the subject opt944 optiflow + adult nasal.Conclusion: we are unable to determine the cause of the reported event.However, it is likely that this was caused by the tube of the opt944 optiflow + adult nasal cannula being pulled.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discoloration and stretching or deformation.Any product that fails the visual inspection is rejected.The subject opt944 optiflow + adult nasal would have met the specification at the time of production.The user instructions which accompany the opt944 optiflow + adult nasal cannula show in pictorial format the correct placement and fitting of the cannula and also warn: - "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." - "do not crush or stretch tube, to prevent loss of therapy.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW + ADULT NASAL CANNULA
Type of Device
BTT
MDR Report Key12458047
MDR Text Key272213568
Report Number9611451-2021-01065
Device Sequence Number1
Product Code BTT
UDI-Device Identifier09420012429117
UDI-Public(01)09420012429117(10)2101085608(11)200422
Combination Product (y/n)N
PMA/PMN Number
K162553
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 08/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT944
Device Catalogue NumberOPT944
Device Lot Number2101085608
Was Device Available for Evaluation? No
Date Manufacturer Received09/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/22/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-