• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION POLYFLUX 140H; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION POLYFLUX 140H; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 112465
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/19/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone no.: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during priming with a polyflux 140h, the product was observed to be broken resulting in a saline leak.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information added to d9, h3, h6 and h10 h10.The device was received for evaluation.The visual inspection by naked eye of the product showed the product wet.A leak test of the dialysate side was performed and a leak was observed due to a crack in the welding seam.The reported condition was verified.The cause is manufacturing related.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYFLUX 140H
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key12461296
MDR Text Key271083867
Report Number9611369-2021-00234
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414102753
UDI-Public(01)07332414102753
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2024
Device Catalogue Number112465
Device Lot Number1-5811-H-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2021
Date Manufacturer Received09/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-