CARDIAC SCIENCE CORPORATION RESPONDER 2000,BASIC,PKG,ENGLISH; AUTOMATED EXTERNAL DEFIBRILLATOR
|
Back to Search Results |
|
Model Number 2025653-001 |
Device Problem
Activation Problem (4042)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device did not respond to the front panel controls.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to a zoll approved service provider.The customer's report was observed and attributed to the power control board.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|