HUDSON REAMING ATTACHMENT FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number 05.001.218 |
Device Problems
Device-Device Incompatibility (2919); Premature Separation (4045)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Concomitant med products and therapy dates: monobloc flexible reamer device ((b)(6) 2021) reporter's phone number was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi : (b)(4).
|
|
Event Description
|
It was reported from (b)(6) that the reaming attachment device and the monobloc flexible reamer device could not be connected well.It was reported that the shape of the attachment connection seemed to be correct, but it did not lock and came off when pulled.The surgeon commented that it could not be judged whether it was a defect of the individual device or a problem with the structure of the product.It was not reported if the event occurred during a surgical procedure.It was not reported if there was a delay to a planned procedure.It was not reported if there was a spare device available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: this device was returned for evaluation; however, during pre-repair assessment, it was determined that the device passed all operational specifications and no failure was identified.Therefore, the reported condition was not confirmed and an assignable root cause was not determined.A review of the device history was performed and no non-conformances were detected related to the reported condition.
|
|
Search Alerts/Recalls
|
|
|