MAUDE Adverse Event Report: ZIMMER BIOMET, INC. M2A-MAGNUM MOD HD SZ 46MM; PROSTHETIC, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. M2A-MAGNUM MOD HD SZ 46MM; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Device Problem Material Erosion (1214)
Patient Problem Metal Related Pathology (4530)
Event Date 05/13/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant products: item # 139256 item name m2a-magnum 42-50 tprinsrt std lot #929850.Item # 51-103090 item name tprlc 133 type1 pps so9x137mm lot # 2588521.The device will not be returned for analysis location unknown; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2021 - 02623.
 
Event Description
It was reported the primary left tha was performed.Subsequently developed elevated metal ions, and pseudotumor.Revision of left tha was performed nine(9) years post implantation.And surgeon noted metallosis of left hip.Head/neck exchange to dual active mobility system.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.  review of the device history records identified no deviations or anomalies during manufacturing no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: patient was revised due to elevated metal ions and pseudotumor.Green staining, pseudotumor removed.Removal of 46mm ball with standard neck, no corrosion noted.Dual mobility bearing placed.A definitive root cause could not be determined if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M2A-MAGNUM MOD HD SZ 46MM
Type of Device
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12467413
MDR Text Key271262150
Report Number0001825034-2021-02624
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2020
Device Model NumberN/A
Device Catalogue Number157446
Device Lot Number990520
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/18/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight64 KG
-
-