A customer reported reflux occurred with a whitish deposit in the eye during a vitreoretinal procedure.Additional information was received from the surgeon who stated small white floating particles were noticed coming out of the vitrectome head during cutting while performing a pars plana vitrectomy procedure, for epi-retinal membrane and multiple retinal tears.The vitrectome was replaced and the procedure was completed; however, postoperatively, surgeon observed the particles at the gas-liquid interface in a patients right eye.
|
A sample was not received at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the product was processed and released according to the product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown; therefore, specific action with regards to this complaint cannot be taken.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|