• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT380
Device Problems Leak/Splash (1354); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).We are currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
 
Event Description
A healthcare facility in (b)(6) reported that a rt380 adult dual-heated evaqua2 breathing circuit failed the leak test prior to patient use.There was no patient involvement.
 
Event Description
A healthcare facility in texas reported that a rt380 adult dual-heated evaqua2 breathing circuit failed the leak test prior to patient use.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).Method: the complaint rt380 adult evaqua2 breathing circuit was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information provided by the customer and our knowledge of the product.Results: the customer reported that a rt380 adult dual-heated evaqua2 breathing circuit failed the leak test prior to patient use.Conclusion: without the complaint device, we are unable to determine what may have caused the reported event.All rt380 adult evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The subject breathing circuit would have met the required specifications at the time of production.Our user instructions that accompany the rt380 adult evaqua2 breathing circuit show in pictorial format the correct set-up of the circuit and also states the following: "check all connections are tight before use." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "set appropriate ventilator alarms.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
MDR Report Key12471548
MDR Text Key272456900
Report Number9611451-2021-01069
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
PMA/PMN Number
K122432
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT380
Device Catalogue NumberRT380
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? No
Date Manufacturer Received10/06/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-