SMITH & NEPHEW, INC. INTERTAN 3.2MM GUIIDE PIN SLEEVE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71674532
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/24/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that, during an internal fixation, an intertan 3.2mm guide pin sleeve would not slide in intertan sleeve.Surgery was completed with a surgical delay greater than 30 min, using a smith and nephew back up device.Patient was not harmed as consequence of this problem.
 
Manufacturer Narrative
H3, h6: the device, used treatment, was returned for evaluation.A visual inspection of the returned devices confirms normal signs of wear and usage.There are some scratches, burrs and gouges on the device.A functional evaluation was conducted and the drill sleeve was not able to target correctly with the mating device confirming the stated failure.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 3.2MM GUIIDE PIN SLEEVE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12472142
MDR Text Key271452623
Report Number1020279-2021-06868
Device Sequence Number1
Product Code HSB
UDI-Device Identifier00885556389348
UDI-Public00885556389348
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674532
Device Catalogue Number71674532
Device Lot Number19MM09182
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2021
Date Manufacturer Received09/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-