• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DURATA STS OPTIM ACTIVE FIXATION; DEFIBRILLATION LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DURATA STS OPTIM ACTIVE FIXATION; DEFIBRILLATION LEAD Back to Search Results
Model Number 7120/65
Device Problems Break (1069); Low impedance (2285); Charging Problem (2892)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/31/2021
Event Type  Injury  
Event Description
During a clinic follow-up, an alert for over current detection (ocd) and a decrease in the defibrillation impedance were noted on the right ventricular (rv) lead.Rv lead damage was suspected, but was unable to be confirmed.The rv lead was capped and replaced to resolve the event.The patient was stable and will continue to be monitored.
 
Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION
Type of Device
DEFIBRILLATION LEAD
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12476148
MDR Text Key271528655
Report Number2017865-2021-30023
Device Sequence Number1
Product Code LWS
UDI-Device Identifier05414734502092
UDI-Public05414734502092
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2014
Device Model Number7120/65
Device Catalogue Number7120-65
Device Lot Number3384238
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/04/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/27/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-