Model Number X SERIES |
Device Problem
No Display/Image (1183)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that while attempting to monitor a patient (age & gender unknown), the device's screen turned completely white and was illegible.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
The device was received at zoll medical corporation.The customer's report was duplicated and attributed to a faulty lcd display module.The lcd module was replaced to remedy the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|