MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. VANTA; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. VANTA; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 977006
Device Problems High impedance (1291); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Manufacturer Narrative
Continuation of d10: product id: 39565-30, serial# (b)(4), implanted: (b)(6) 2013, product type: lead, product id: 39565-30, serial# (b)(4), implanted: (b)(6) 2013, product type: lead.Other relevant device(s) are: product id: 39565-30, serial/lot #: (b)(4), ubd: 20-sep-2016, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient via manufacturer representative who was implanted with an implantable neurostimulator (ins) for non-malignant pain.It was reported that the patient was seen for post-op on (b)(6) 2021, and impedances were over 40,000 ohms on the 8-15 side of the paddle.Only 11 was okay.Reprogramming was attempted to try and offer him right back and leg relief despite the impedances.The patient will report back if they are getting good coverage, or will take the next steps for a possible lead revision.The patient didn't report any notable fall or accident which may have caused the issue.The patient is getting some relief.At this time, the health care professional has no further information regarding the event.Surgical intervention was not performed at this time; however, the patient was referred to their neurosurgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VANTA
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key12478909
MDR Text Key271672154
Report Number3004209178-2021-13920
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00763000411923
UDI-Public00763000411923
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2023
Device Model Number977006
Device Catalogue Number977006
Was Device Available for Evaluation? No
Date Manufacturer Received08/26/2021
Date Device Manufactured04/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-