DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Erratic or Intermittent Display (1182)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device (10/213): a getinge field service engineer (fse) evaluated the iabp and was unable to duplicate reported issue.Powered the unit on and off 6 times, ran on patient simulator for 2 hours while flexing coiled cable and articulating monitor with no issues found.Functional testing and safety check to factory specifications.Unit passed all functional and safety tests per factory specifications.Returned to customer and cleared for clinical use.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during a training the cardiosave intra-aortic balloon pump (iabp) had the display flickering during bootup.There was no patient involvement, and no adverse event reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, g4, g7, h2, h4, h6(investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: g1(contact person).
|
|