• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE BD SYRINGE 0.3ML 30GA 8MM; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE BD SYRINGE 0.3ML 30GA 8MM; PISTON SYRINGE Back to Search Results
Catalog Number 326638
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2021
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone# : (b)(6).Initial reporter zip code : (b)(6).Investigation summary : exec summary - no samples (including photos) were returned therefore bd was not able to duplicate or confirm the customer¿s indicated failure and the root cause is undetermined.A review of the complaint lot history check was performed and this is the 4th.Related complaint for hub separates on this lot #.No non-conformances were raised in association with this type of event for this lot concluding all inspections were performed as per the applicable operations and met qc specifications.Capa/sa - based on the above no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) are required at this time.Dhr review - a lot history review was carried out and no related non conformances were raised in association with this packaged lot concluding all inspections were performed as per the applicable operations and met qc specifications.
 
Event Description
It was reported that 2 bd syringe 0.3ml 30ga 8mm hubs separated from the device.The following information was provided by the initial reporter : the customer reported about the needle and shield separation from the barrel when removing the shield.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SYRINGE 0.3ML 30GA 8MM
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12481626
MDR Text Key271702484
Report Number1920898-2021-00997
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K024112
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor,foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number326638
Device Lot Number0286309
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/19/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/12/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-