• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97715
Device Problems Low impedance (2285); Patient Device Interaction Problem (4001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/26/2021
Event Type  malfunction  
Manufacturer Narrative
Concomitant products: product id: 977a2, serial#: unknown, implanted: (b)(6) 2018, product type: lead.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The healthcare professional (hcp) reported through a manufacturer representative that electrodes 10, 13, and 15 had impedances displayed as ¿avoidance¿ and that in some cases the values ¿returned to the normal display when the patient¿s posture was changed.¿ the implantable neurostimulator (ins) session report from (b)(6) 2021 noted that electrodes 10, 13, and 15 had low impedances of 90, 90, and 200 respectively.It was stated that ¿the lead position was not at the contact position¿ and that this may have led or contributed to the issue.Clarification was sought regarding this statement; follow-up received stated that it was meant that since the three electrodes in question were on the same lead that they ¿were not in contact with each other¿ and since the patient¿s two leads were not implanted in parallel, ¿it was difficult to think that the electrodes contacted each other.¿ the patient¿s physician¿s observation about causality was unknown.It was noted that the issue was considered resolved at the time of report.There were no surgical interventions planned or performed and the intractable chronic pain patient was alive with no injury at the time of report.No complications were reported or anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key12482253
MDR Text Key271728193
Report Number3004209178-2021-13962
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2019
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/26/2021
Date Device Manufactured08/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-