Model Number 0998-00-0800-53
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/23/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and was able to duplicate the reported issue.The fse observed that power was present in the cart.When inspecting the console, the fse found saline damage on the power management board.The fse resolved the reported issue by replacing the power management board.Subsequently, the fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) experienced a power failure.The iabp unit was swapped out with another iabp unit to continue therapy without issue.There was no patient harm and no adverse event report.Patient height: 180.3 cm.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields - b4, e1 (site country), g3, g6, g7, h2, h4, h6 (type of investigation, investigation findings, investigation conclusions), h10.A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and observed power present in the cart.When inspecting the console, the fse found saline damage on the power management pcb.The fse resolved the reported issue by replacing the power management pcb (0670-00-1162).The fse performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12484816
MDR Text Key271907049
Report Number2249723-2021-02111
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
ASKU (ASKED BUT UNKNOWN)
Patient Age56 YR
Patient SexMale
Patient Weight95 KG
-
-