DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/23/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and was able to duplicate the reported issue.The fse observed that power was present in the cart.When inspecting the console, the fse found saline damage on the power management board.The fse resolved the reported issue by replacing the power management board.Subsequently, the fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) experienced a power failure.The iabp unit was swapped out with another iabp unit to continue therapy without issue.There was no patient harm and no adverse event report.Patient height: 180.3 cm.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields - b4, e1 (site country), g3, g6, g7, h2, h4, h6 (type of investigation, investigation findings, investigation conclusions), h10.A getinge field service engineer (fse) was dispatched to investigate.The fse evaluated the iabp unit and observed power present in the cart.When inspecting the console, the fse found saline damage on the power management pcb.The fse resolved the reported issue by replacing the power management pcb (0670-00-1162).The fse performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|