ABBOTT QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3562
Device Problem Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/03/2021
Event Type  malfunction  
Event Description
During an in clinic follow up following a mri, it was noted that the lv capconfirm test was unable to be performed; however when manually tested, the lv threshold was stable.A fixed output was programmed to resolve the event.The patient was stable.
 
Manufacturer Narrative
The complaint of being unable to complete a lvcap confirm threshold test after an mri was confirmed, consistent with fusion occurring when there is a transition from capture to loss of capture.The device is performing per specifications and there is no indication that the observed behavior was related to the mri.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12487688
MDR Text Key271881922
Report Number2017865-2021-30718
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2021
Device Model NumberPM3562
Device Catalogue NumberPM3562
Device Lot NumberP000085665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-