DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,INTL SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-UC-3013-55 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/26/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device (10/213): a getinge field service engineer (fse) evaluated the iabp unit for the reported "error# 35" and the found that the main board to the motor controller was defective.
To fix the issue, the fse replaced the main board to motor controller pcb cable (part no- 0012-00-0759-02), and performed all functional and safety checks to meet factory specifications.
Unit passed all functional and safety test per factory specifications.
The iabp was then released to the customer and cleared for clinical service.
A supplemental report will be submitted upon completion of our investigation.
(b)(6).
|
|
Event Description
|
It was reported that during a routine check the cs100 intra-aortic balloon pump (iabp) was showing an electrical test failure code #53 intermittently.
There was no patient involvement, and no adverse event reported.
|
|