• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. AC3 OPTIMUS IABP NA/EMEA SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. AC3 OPTIMUS IABP NA/EMEA SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN001112
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/20/2021
Event Type  malfunction  
Manufacturer Narrative
Qn# (b)(4).
 
Event Description
It was reported that there was an "error in the functionality of the touchscreen+ purge failure. " as a result, the pump was swapped out. There was no report of patient complications, serious injury or death. The pcs (pneumatic control system) and the display head assy were replaced, and the pump was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAC3 OPTIMUS IABP NA/EMEA
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kaylia pen
16 elizabeth drive
chelmsford, MA 01824
MDR Report Key12489503
MDR Text Key271985624
Report Number3010532612-2021-00259
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 08/20/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberIPN001112
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/08/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-