DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/26/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to the customer sight.The fse evaluated the iabp unit and discovered slot 1 will register a battery and was capable of charging.Issue followed the battery to slot 2.Both batteries were replaced with no further issues.The fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.The full name of the event site in block e1 was shortened due to field character limit; the full name is (b)(6).
|
|
Event Description
|
It was reported that the battery in slot one of the cardiosave intra-aortic balloon pump (iabp) would not charge.It is unknown under which circumstances this event occurred; however there was no patient involvement, and no adverse event was reported.
|
|
Event Description
|
N/a.
|
|