CARDIAC SCIENCE CORPORATION G3 PH, AHA 2005, PKG, US ENGLISH; AUTOMATED EXTERNAL DEFIBRILLATOR
|
Back to Search Results |
|
Model Number 9300E-501 |
Device Problem
High impedance (1291)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during functional testing, the device failed for high impedance.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical corporation; the customer's report was duplicated and attributed to a faulty battery terminal 1 on the main board.The main board was replaced to resolve the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|