Model Number X SERIES |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device was unable to obtain an ecg signal via the multi-function cable.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
The device was not returned to zoll medical corporation for evaluation.Instead, a zoll field representative evaluated the device at the customer's site.The customer's report was observed during initial testing.The multifunction cable was returned to zoll medical corporation and the customer's report was unable to be duplicated.The multifunction cable passed continuity testing and was able to properly connect to a test cpr adapter.The cable was tested with a test device using the test cpr adapter and pads without duplicating the report.The multifunction cable was scrapped at zoll.Analysis of reports of this type has not identified an increase in trend.
|
|