This is filed to report the device was used past the expiration date.It was reported that the mitraclip procedure was performed on (b)(6) /2021 to treat functional mitral regurgitation with an mr grade of 4+.The user was informed that the steerable guide catheter (sgc) was expired.The expiration date for the sgc was 7/23/2021.Despite knowing the sgc was expired, the sgc was still used for the procedure.Two clips were implanted, reducing mr to 2.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.It should be noted that per the instructions for use states, mitraclip g4 system, ce ¿do not use the mitraclip g4 system after the ¿use by¿ date stated on the package label, and never reuse or re-sterilize the system.Use of expired, reused, or re-sterilized devices may result in infection¿.In this case, it was reported that the user used the device despite knowing the sgc was expired which is a deviation from instruction for use.The investigation determined the reported improper or incorrect procedure or method (use after expiration) was due to the user error of using the device despite knowing the steerable guide catheter/sgc was expired.There is no indication of a product issue with respect to manufacture, design or labeling.
|