• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR Back to Search Results
Catalog Number SGC0302
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/27/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report the device was used past the expiration date.It was reported that the mitraclip procedure was performed on (b)(6) /2021 to treat functional mitral regurgitation with an mr grade of 4+.The user was informed that the steerable guide catheter (sgc) was expired.The expiration date for the sgc was 7/23/2021.Despite knowing the sgc was expired, the sgc was still used for the procedure.Two clips were implanted, reducing mr to 2.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints reported from this lot.It should be noted that per the instructions for use states, mitraclip g4 system, ce ¿do not use the mitraclip g4 system after the ¿use by¿ date stated on the package label, and never reuse or re-sterilize the system.Use of expired, reused, or re-sterilized devices may result in infection¿.In this case, it was reported that the user used the device despite knowing the sgc was expired which is a deviation from instruction for use.The investigation determined the reported improper or incorrect procedure or method (use after expiration) was due to the user error of using the device despite knowing the steerable guide catheter/sgc was expired.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key12500068
MDR Text Key273631788
Report Number2024168-2021-08322
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/23/2021
Device Catalogue NumberSGC0302
Device Lot Number00723U235
Was Device Available for Evaluation? No
Date Manufacturer Received10/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-