Model Number 1804350-08
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2021
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported material rupture; however, factors that may contribute to material ruptures include, but are not limited to, material damage, interactions with other devices or interaction with lesion calcification and tortuosity.Additionally, there was no damage or leak noted to the stent delivery system (sds) during the inspection prior to use or during preparation of the device, which suggests a product quality issue did not contribute to the reported difficulties.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that the procedure was to treat a heavily calcified, moderately tortuous, 90% stenosed lesion in the mid left anterior descending coronary artery (mlad).A 3.5x08mm xience skypoint drug eluting stent (des) was advanced to the lesion and inflated but the balloon ruptured on the first inflation to 12 atmospheres (atm).The des was removed and a new, same size xience skypoint des was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SKYPOINT DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key12503068
MDR Text Key272946833
Report Number2024168-2021-08362
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648233371
UDI-Public08717648233371
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 09/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/13/2023
Device Model Number1804350-08
Device Catalogue Number1804350-08
Device Lot Number0090741
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-