Catalog Number 302100
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/16/2021
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: unknown.Device manufacture date: unknown.Investigation summary: one photo and one sample were received by our quality team for evaluation.From the photo, it was observed that the plunger rod detaches from the stopper and the plunger head is missing.From the returned sample, the plunger was observed to be separated from the stopper due to plunger head chip off.A device history record could not be evaluated as the lot number is unknown.The probable root cause of the plunger head chip off could have be due to the molded plunger tip hitting against the molded plunger target box when transferring from the molding machine to the assembly line.A curtain has been installed at the molded plunger target box in order to act as a cushion and reduce impulse.Communication with the production technicians to raise awareness of the reported defect will occur.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that the bd¿ tuberculin syringe stopper separated from the plunger.The following information was provided by the initial reporter, translated from (b)(6) to english: "the customer uses this product for covid-19 vaccination.The customer reported that the stopper was separated from the plunger.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD TUBERCULIN SYRINGE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12508766
MDR Text Key272695583
Report Number8041187-2021-00853
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number302100
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/25/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-