Model Number X SERIES
Device Problem Unable to Obtain Readings (1516)
Patient Problem Necrosis (1971)
Event Date 08/30/2021
Event Type  Death  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that while attempting to defibrillate a female patient (age unknown), the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that the patient subsequently expired, however it was not a result of the reported malfunction.
 
Manufacturer Narrative
The device was returned to zoll medical corporation.The customer's report was duplicated and attributed to a missing gasket in the multifunction cable.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key12509162
MDR Text Key272561556
Report Number1220908-2021-03017
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946004354
UDI-Public00847946004354
Combination Product (y/n)N
PMA/PMN Number
K112432/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 08/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2021
Was the Report Sent to FDA? No
Date Manufacturer Received08/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Death;
-
-