• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Device Problem Failure to Charge (1085)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device failed to charge.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device passed full functional testing including defib cycling without duplicating the report.An internal inspection of the device found no discrepancies.The device was recertified and returned to the customer.The battery was not returned as part of the investigation.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key12509299
MDR Text Key272563109
Report Number1220908-2021-03023
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946019143
UDI-Public00847946019143
Combination Product (y/n)N
PMA/PMN Number
K112432/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 08/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2021
Was the Report Sent to FDA? Yes
Event Location Other
Date Manufacturer Received08/31/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-