Model Number 0998-00-0800-53 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/30/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to the customer sight.The fse evaluated the iabp unit and could not duplicate the issue.Device was plugged in and fully charged.Fse disconnected device and plugged back in multiple times with no issues.Nothing out of ordinary in logs.The fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.
|
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) battery was being plugged in but charge level was not increasing.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
|
|
Event Description
|
It was reported that during use the cardiosave intra-aortic balloon pump (iabp) battery was being plugged in but charge level was not increasing.No patient harm, serious injury or adverse event was reported.
|
|
Manufacturer Narrative
|
Customer contact information was provided for (b)(6) and is as follows: (b)(6).Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period sep 2019 through aug 2021 was reviewed.There were no triggers identified for the review period.
|
|
Event Description
|
N/a.
|
|