Model Number 3660
Device Problem Wireless Communication Problem (3283)
Patient Problem Inadequate Pain Relief (2388)
Event Date 08/20/2021
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.
 
Event Description
It was reported the patient underwent a non-related surgical procedure wherein it is unknown if the ipg was placed into surgery mode prior to the procedure. As a result, the patient's ipg was unable to communicate with external devices and the patient lost therapy. A manufacturer representative confirmed the issue and the ipg was deemed inoperable. Surgical intervention may occur at a later date to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR
Type of DeviceSCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key12510408
MDR Text Key272657325
Report Number3006705815-2021-04702
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date09/11/2021
Device Model Number3660
Device Catalogue Number3660
Device Lot NumberA000086512
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received10/01/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 09/21/2021 Patient Sequence Number: 1
-
-