Catalog Number UNKNOWN
Device Problem Structural Problem (2506)
Patient Problem Perforation of Vessels (2135)
Event Type  Injury  
Manufacturer Narrative
Initial reporter occupation: non-healthcare professional.Investigation: the following allegations have been investigated: vena cava perforation.The reported allegations have been investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Catalog number and lot number are unknown.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
 
Event Description
The following information is alleged: the patient received a celect inferior vena cava (ivc) filter on (b)(6) 2014 and multiple prongs of the filter have perforated the ivc wall.Hospital and medical records have been requested, but not yet provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
Manufacturer Contact
lissi walmann
sandet 6
bjaeverskov 4632 
56868686
MDR Report Key12517011
MDR Text Key272907759
Report Number3002808486-2021-01817
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Other
Type of Report Initial
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received09/01/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-