It was reported that the arctic sun device was failing calibration for an error 80 and error 1 (ran it multiple times).
It was clear the circulation pump was intermittently failing and the device was not heating correctly.
The user requested a quote for the 2000 hour service package.
Per evaluation, the arctic sun device was failing calibration due to low flow.
Per sample evaluation, it was reported that the tank gaskets are worn and torn.
It stated that the main voltage circuit card spade terminal showed signs of electrical stress.
The power inlet module and main voltage circuit card were replaced due to the electrical stress.
|