Model Number 3660 |
Device Problem
Battery Problem (2885)
|
Patient Problems
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event is estimated.
|
|
Event Description
|
It was reported that diagnostics indicated the elective replacement indicator (eri) triggered earlier than intended.The device has the appropriate level of battery voltage to provide therapy.
|
|
Manufacturer Narrative
|
A system displaying warning message ¿replace generator soon¿ warning message was reported to abbott.As a result, the patient¿s ipg was explanted and replaced.Therapy was restored.The results of the investigation are inconclusive since the implant was not returned for analysis.Based on the information received, the reported event is consistent with the implant reaching end of service.
|
|
Event Description
|
Additional information indicated that a surgical intervention occurred wherein the ipg was explanted and replaced to address the issue.
|
|