Model Number 3660
Device Problem Battery Problem (2885)
Patient Problems Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/08/2021
Event Type  malfunction  
Manufacturer Narrative
Date of event is estimated.
 
Event Description
It was reported that diagnostics indicated the elective replacement indicator (eri) triggered earlier than intended.The device has the appropriate level of battery voltage to provide therapy.
 
Manufacturer Narrative
A system displaying warning message ¿replace generator soon¿ warning message was reported to abbott.As a result, the patient¿s ipg was explanted and replaced.Therapy was restored.The results of the investigation are inconclusive since the implant was not returned for analysis.Based on the information received, the reported event is consistent with the implant reaching end of service.
 
Event Description
Additional information indicated that a surgical intervention occurred wherein the ipg was explanted and replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
MDR Report Key12523805
MDR Text Key273123986
Report Number3006705815-2021-04708
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020192
UDI-Public05415067020192
Combination Product (y/n)N
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 10/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/04/2021
Device Model Number3660
Device Catalogue Number3660
Device Lot NumberA000084342
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-