Intuitive surgical, inc.(isi) received the maryland bipolar forceps instrument involved with this complaint and completed the device evaluation.Failure analysis investigations replicated/confirmed the customer reported complaint.The instrument was found to have a broken conductor wire at the yaw pulley on the distal end.The conductor wire was broken at the grip-crimp location.The instrument failed the electrical continuity test thermal damage was observed near the conductor wire-yaw pulley entry.The housing was removed from the back end, and no damage was found.The root cause for the broken conductor wire is attributed to a component failure.Additional observations that were not reported by the site were also identified.The instrument was found to have damage to the conductor wire¿s insulation.The insulation exhibited thermal damage.The insulation exhibited signs of melting, however, no pieces were missing.The conductor wires were exposed, however no wires were damaged.The root cause of the damaged conductor insulation is attributed to a component failure.The instrument was found to have thermal damage on the bipolar yaw pulley.The yaw pulley showed signs of charring and localized melting the base and on the exterior of the grips.The thermal damage to the yaw pulley is indicative of arcing.The damage measured 0.025" x 0.070" in length.The root cause for this failure is attributed to a component failure.A review of the site's complaint history does not reveal any related complaints involving this product or this event.A review of system logs for system (b)(4) with a procedure date of (b)(6) 2021, confirmed a maryland bipolar forceps (pn# 470172 - 16 / lot # n10201006 - 0210) was exchanged with another maryland bipolar forceps (pn# 470172 - 16 / lot # n10201006 - 0068) during this procedure.The instrument has a maximum of 10 uses.The alleged event occurred on the 4th use of the instrument.There were 6 more uses remaining.Based on the additional information obtained from failure analysis investigations, this complaint is being reported due to the following conclusion: the instrument had conductor wire damage with no evidence of user mishandling or misuse.Thermal damage at or near a conductor wire breakage is evidence of electrical discharge at a location other than intended.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.
|