DATASCOPE CORP. - FAIRFIELD TRANS-RAY PLUS 7.5 FR. 35CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0684-00-0604
Device Problems Difficult to Flush (1251); Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event Description
It was reported that after inserting the intra-aortic balloon (iab) the arterial pressure signal could not be obtained from the lumen.The customer attempted to flush but met resistance.The iab was removed and replaced with a new one.There was no patient harm or adverse event reported.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period sep-19 through aug-21 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #(b)(4).
 
Manufacturer Narrative
Updated fields: b4, d10, g4, h3, h6, h10.The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id # (b)(4).
 
Event Description
N/a.
 
Event Description
N/a.
 
Manufacturer Narrative
The product was returned with the membrane completely unfolded and blood on the exterior of the catheter.The extender and pressure tubing was also returned.A kink was found on the catheter tubing near the y-fitting approximately 72.4cm from the iab tip.Additionally a second kink was found on the catheter tubing and inner lumen approximately 73.4cm from the iab tip.The technician attempted to flush/aspirate the inner lumen and was unable to do so.The technician then attempted to insert a 0.025¿ laboratory guide wire through the inner lumen and resistance was only felt at the kinked location of 73.4cm.A sensor output test was performed and the sensor was found to be within specification.An underwater leak test of the balloon, catheter, y-fitting, extracorporeal, extender and pressure tubing was performed and no leaks were detected.A kink in the inner lumen can cause difficulty flushing the inner lumen the ability to pressure monitor.We are unable to determine when the kink may have occurred.The evaluation confirmed the reported problems.A lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANS-RAY PLUS 7.5 FR. 35CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key12532994
MDR Text Key273341688
Report Number2248146-2021-00627
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/18/2022
Device Catalogue Number0684-00-0604
Device Lot Number3000109191
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-