Catalog Number 0684-00-0604 |
Device Problems
Difficult to Flush (1251); Pressure Problem (3012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
|
|
Event Description
|
It was reported that after inserting the intra-aortic balloon (iab) the arterial pressure signal could not be obtained from the lumen.The customer attempted to flush but met resistance.The iab was removed and replaced with a new one.There was no patient harm or adverse event reported.
|
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period sep-19 through aug-21 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #(b)(4).
|
|
Manufacturer Narrative
|
Updated fields: b4, d10, g4, h3, h6, h10.The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id # (b)(4).
|
|
Event Description
|
N/a.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
The product was returned with the membrane completely unfolded and blood on the exterior of the catheter.The extender and pressure tubing was also returned.A kink was found on the catheter tubing near the y-fitting approximately 72.4cm from the iab tip.Additionally a second kink was found on the catheter tubing and inner lumen approximately 73.4cm from the iab tip.The technician attempted to flush/aspirate the inner lumen and was unable to do so.The technician then attempted to insert a 0.025¿ laboratory guide wire through the inner lumen and resistance was only felt at the kinked location of 73.4cm.A sensor output test was performed and the sensor was found to be within specification.An underwater leak test of the balloon, catheter, y-fitting, extracorporeal, extender and pressure tubing was performed and no leaks were detected.A kink in the inner lumen can cause difficulty flushing the inner lumen the ability to pressure monitor.We are unable to determine when the kink may have occurred.The evaluation confirmed the reported problems.A lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint #(b)(4).
|
|