• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Appropriate Term/Code Not Available (3191)
Patient Problems Bacterial Infection (1735); Stroke/CVA (1770); Fall (1848); Laceration(s) (1946); Renal Failure (2041); Sepsis (2067); Transient Ischemic Attack (2109); Hypovolemia (2243); Hypervolemia (2664); Unspecified Vascular Problem (4441); Heart Failure/Congestive Heart Failure (4446); Gastrointestinal Hemorrhage (4476); Renal Impairment (4499)
Event Date 07/23/2020
Event Type  Injury  
Manufacturer Narrative
Medtronic was made aware of this event through a search of literature publications.It was not possible to ascertain specific de vice information from the literature publication or to match the event with previously reported events.This information is based entirely on journal literature.All information provided is included in this report.Patient information is limited due to confidentiality concerns.Multiple patients and multiple manufacturers were noted in the article; however, a one-to-one correlation could not be made with unique device serial numbers.The baseline gender/age of the patients represented in the article is male/55 years old.Without a lot number or device serial number, the manufacturing date cannot be determined.Since no device id was provided, it is unknown if this event has been previously reported.Request for additional information will be made and, upon receipt, a supplemental report will be submitted accordingly.Referenced article: outcomes among patients with left ventricular assist devices receiving maintenance outpatient hemodialysis: a case series.American journal of kidney disease, february 2021; 77(2):226-234.Doi: 10.1053/j.Ajkd.2020.04.018.Pmid: 32711070 additional information has been requested regarding the cause of the events, device serial numbers and patient demographic data, but it was not available at the time of this report.If additional information is received, the event will be updated and a supplemental report will be sent.Investigation of this event is pending and a supplemental report will be sent upon its completion.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A journal article was reviewed which contained information regarding ventricular assist devices (vads).The article discussed outcomes after maintenance dialysis for vad patients with kidney failure.Multiple patients and multiple manufacturers were noted in the article; however, a one-to-one correlation could not be made with unique device serial numbers.There were patients on vad support who experienced acute kidney injury after vad implantation, requiring maintenance hemodialysis.Patients on vad support also required hospitalization for mechanical falls with lacerations, arterial spasm, improper anticoagulation status, bacterial infections/bacteremia, driveline infections, hypervolemia/hypovolemia, heart failure exacerbations, gastrointestinal bleeding and transient ischemic attack (tia)/cerebrovascular accidents (cvas).The vads remain in use.No further patient complications have been reported as a result of this event.
 
Event Description
It was reported that dialysis sessions had to be stopped prematurely due to one instance of an uncharacterized vad alarm.
 
Manufacturer Narrative
A supplemental report is being submitted for corrections.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
A supplemental report is being submitted for investigation completion.Product event summary: pumps with unknown serial numbers were not returned for evaluation.Review of the sterility certificates could not be conducted since the serial numbers are unknown.Log file analysis was not conducted since log files were not available.There is insufficient information regarding the reported "uncharacterized vad alarm" event; as a result, that reported event could not be confirmed.There is no evidence to suggest a device malfunction caused or contributed to the reported events.Possible clinical factors that may have contributed to these events include the patients¿ pre-existing histories and related comorbidities, the progression of their underlying diseases, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patients¿ complex post-operative courses.There are possible patient, pharmacological and procedural factors that may have contributed to these events.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key12533161
MDR Text Key273374605
Report Number3007042319-2021-06431
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Literature,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Hospitalization;
Patient Age55 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-